Installation and Operation Manual
X-VA-1110-1140-eng
Part Number: 541B040AAG
September, 2012
Plastic protective sleeve must remain over glass
tube. (Meter sizes 7 -13 only)
Fasten meter windows securely.
Do not operate above pressure and temperature
limits.
Avoid pressure and flow surges.
Do not service or repair while pressurized.
Read and understand instruction manual.
Failure to comply could result in serious personal
injury or property damage.
Table 1-1 Pressure Ratings and PED Category
End
Maximum Operating Pressure (psig)
Meter
Fitting
@ 70°F
Size
Material
(21°C)
2
CPVC
100
Metal
500
316 SS
200
(w/CRN)
6
CPVC
100
Metal
450
316 SS
200
(w/CRN)
7
CPVC
100
Metal
350
8
CPVC
100
Metal
300
316 SS
160
(w/CRN)
9
PVC
100
Metal
175
316 SS
100
(w/CRN)
10
PVC
100
Metal
100
316 SS
60
(w/CRN)
12
Metal
75
13
Metal
100
NOTES:
1
Maximum temperature rating for nonmetal end fittings is 150°F
(66°C) for all materials except PVC, which is 140°F.
2
Flanged unit pressure ratings are limited to ANSI flange or tube ratings
whichever is lower.
3
Sizes 12 and 13 do not conform to Pressure Equipment Directive
97/23/EC, therefore it cannot be sold or used in the EU/EFTA.
Section 1 Introduction
®
Brooks
1110 and 1140 Series
WARNING
GLASS TUBE
EXPLOSION
HAZARD
@ 150°F
1
@ 250°F
PED
(66°C)
(121°C)
CAT.
100
Not Available
500
500
200
200
100
Not Available
450
450
200
200
100
Not Available
350
350
SEP
100
Not Available
300
300
160
160
100 (Note 1) Not Available
175
175
100
100
100 (Note 1) Not Available
100
100
60
60
75
75
SEE
NOTE
100
100
2
3
1-3