Backlight/Laser Pointer
Pulling the trigger automatically turns the backlight on for 8 seconds and also
illuminates the laser pointer.
NOTE: During measurement, it is best to ensure that the measured target
diameter is twice the spot size (S) of the thermometer, and then determine the
test distance (D) according to the D:S diagram (refer to D:S part).
For example, if you use the 61-827 to measure the temperature of an object
with a diameter of about 4" (10cm), then according to the above, the spot
size (S) of the thermometer should be about 2" (5cm) for highest accuracy,
and according to the D:S diagram, the measured distance (D) is about 20"
(50 cm).

Auto Power Off

If there is no operation for 8 seconds, the thermometer will automatically
power off and save the currently held measurement.