8. TROUBLESHOOTING
Phenomenon
Upper Line - - - -
Under Line_ _ _ _
Err-1
Err-2
9. PACKING LIST
Description
Balance
Pan
Power cord
Manual
Possible Reason
Overload
Overload or load cell broken
Too frequently turn on and off
the balance
The balance is not stabilized
Low battery
QTTY
1 pcs
1 pcs
1 pcs
1 pcs
17
Solution
Re-calibrate the balance
Re-calibrate the balance
Turn off the balance, resume it after 3
seconds
Wait for a few seconds for stabilization
Charge or replace battery