To Use In Right Way - MetaBiomed e-magic finder Serie Instrucciones De Uso

Tabla de contenido

Publicidad

[Figure 5-3]
5.3 Input Information on a new patient
5.3.1 Please type in the surname, last name and patient identity code. (Patient identity code
can not be designated twice.)
5.3.2 Please store the input information by pressing New Patient button.
5.4 Open restored patient information
5.4.1 Please type in the surname, last name or identity code of a patient. (You are
recommended to use the patient identity code because the information can not be
opened if there are patients who use the same family names or same last names. You
can also open patient information only with a patient identity code.)
5.4.2 You can open patient information with a click of SEARCH button. (It is possible to
move to MEASUREMENT window without the input of the information of a patient;
however, a patient should be selected to save a measurement value. Therefore, patient
information is advised to be selected before a measurement is performed.)
5.5 Measure the working length of the root canal and save it.
5.5.1 Please input new patient information or open stored patient information.
5.5.2 Please click Measurement button.
5.5.3 Please measure the working length of the root canal. (The value displayed on the APEX
program could appear later than that shown on the LCD of the main unit because of the
specification of a PC or peripheral devices on the computer.)
5.5.4 Please input the number of the tooth which is measured, measured working length, and
other remarks to be recorded.
5.5.5 Please click SAVE button to store the input data
5.6 Print out stored measurement values in a hard form.
5.6.1 Please select the information of a patient. [Figure 5-7]
5.6.2 Please click PRINT button to print out the patient information stored in Patient
History .
20
[Figure 5-4]

6. To Use in right way

6.1. Guide for making accurate measurement
6.1.1. Please power on prior to use.
6.1.2. It is advisable measurement way that put a half of file along with the canal first, and then
6.1.3. In case of no indication when the file reached to the APEX, it has possibility that the root
6.1.4. If there is foreign materials like irrigants in the mouth, there has possibility for electric
6.1.5. If necrotic dentin and cementum in the root canal are not removed, there are possibility for
[Figure 5-5]
[Figure 5-6]
[Figure 5-7]
connect file to file holder.
canal would be dried perfectly. It is advisable measurement way that the dried root canal
should be filled with saline.
leakage between file and lip holder through the above foreign materials. It is advisable
measurement way that remove the foreign materials which are in the mouth.
mis-measurement. It is advisable for you to enlarge the root canal with a enough size of
file which can be felt slight resistance (snugg y binding) in side of root canal.
21

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Smarpex

Tabla de contenido