when using a paper/plastic pouch. After warm-up is completed, operate at a sterilizing temperature and
pressure of 134ºC/273ºF to 137ºC/279ºF and pressure of 216kPa/31psi for 3 to 12 minutes.
Alternate method: Place non-bagged adapter and thumb nut into the steam autoclave. After warm-up
is completed, operate at a sterilizing temperature of 134ºC/273ºF to 137ºC/279ºF and pressure of
216kPa /31psi for 3 to 4 minutes. Components should be used immediately.
Drying:
To dry, use the drying cycle of the autoclave. Set cycle for 20-30 minutes.
Maintenance, inspection and testing:
• Visually inspect to ensure all contamination is removed.
• Check for distortion, damage and wear. Discard damaged, worn or corroded components.
Storage:
To maintain sterility, instruments should remain bagged until ready to use.
MANUFACTURER CONTACT
In the United States, contact DENTSPLY Professional Customer Service Technical Support at
800-989-8826. For areas outside the United States, contact your local DENTSPLY Representative.
TROUBLESHOOTING
PROBLEM: Water leaking, wet air, and/or loose tips, indicates either:
a.) Center hole of Sani-Tip disposable tip is not fully seated over water spindle inside new adapter.
b.) Correct Sani-Tip Conversion Kit is not properly installed, and/or missing parts; thumb nut has
not been fully tightened, or incorrect tips are being used.
c.) Internal water spindle is damaged.
SOLUTION:
a.) Loosen thumb nut, when interchanging Sani-Tip disposable tips. Insert new tip and push into
adapter as far as possible. While still pushing inward, twist tip 1/2 turn clockwise to create a
watertight seal over internal tapered water spindle. Tighten thumb nut.
b.) Make sure correct Sani-Tip Conversion Kit is properly installed. Confirm presence of plastic
cone and all o-rings. Fully tighten thumb nut after inserting new tip. Verify tips are clear plastic
a nd manufactured by DENTSPLY Professional for the Sani-Tip
Den-Tal-Ez unit is manufactured after 1/97, syringe will require A-1 Conversion Kit.
c.) If internal water spindle inside Sani-Tip adapter is damaged, replace adapter.
PROBLEM: Tip does not fit through thumb nut.
SOLUTION: Loosen thumb nut a full turn to expand opening of plastic cone.
PROBLEM: Little or no water.
SOLUTION: Check for clogged or damaged water spindle inside adapter, or blocked water outlet in
syringe head.
System. Important: If
®