Over-Limit Indication; Limit Settings - Gima OXY-100 Manual De Uso Y Mantenimiento

Oxímetro
Ocultar thumbs Ver también para OXY-100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Degree of protection against harmful ingress of liquids: Ordinary equipment
without protection against ingress of water.
Electro-Magnetic Compatibility: Group I, Class B

6 Over-limit Indication

6.1 Limit settings

- SpO
low limit setting range: 85% ~ 95%.
2
- Pulse Rate limits setting range:
High: 100bpm--240bpm Low: 30bpm--60bpm
During the measurement, if the measured value exceeds the preset value,
the alert beeping sound will be activated, the value that is over-limit will blink
at the same time.
6.2 Alert sound mute setting
- During the measurement, if the alert sound is set to on, short time press "
" recall/mute key, then the alert sound will mute for 90 seconds, but the
over-limited value still keeps blinking. At this moment, the sound indication
icon " " becomes "
alert sound will be activated again.
- During the measurement, if the probe is off or disconnected, the message
"Check Probe" shows on the display screen. The alert sound starts (interval is
5 seconds) and lasts for about 1 minute. If the probe is still off, the Oximeter
will power off automatically.
7 Accessories
1. A probe
2. A holder
3. Battery "AA" × 3
4. User Manual
5. Quality Inspection Certificate
6. A data cable (optional)
7. Oximeter Data Manager software (optional)
Note: The accessories are subject to change. See the Packing
List for detailed items and quantity.
ENGLISH
". If this alert event persists over 90 seconds, then the
34

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido