Brooks Instrument Ar-Mite MT3750 Manual Del Usuario página 14

Ocultar thumbs Ver también para Ar-Mite MT3750:
Tabla de contenido

Publicidad

Installation and Operation Manual
X-VA-MT3750C-eng
Part Number: 541B063AAG
October, 2017
Figure 1-6 Model MT3750C with
Inductive Alarm
Table 1-9 Certifications - MT3750C With Inductive Alarm
Intrinsically Safe
Mark
Agency
Certification
Zone 1, Zone 21
II 2 G Ex ia IIC T6
ATEX
II 2 D Ex ia D 21 IP65 T75°C
Zone 1, Zone 21
IECEx
Ex ia IIC T6 Gb
Ex ia IIIC T
Class I, II, III, Div.1, Groups A thru G, T6
Class I, Zone 0, Zone 1 AEx ia IIC, T6
CSA
Ex ia IIC T6
Ex ia IIC T6 Gb
NEPSI
(China)
Non-Incendive
Class I, II, III, Div. 2, Groups A thru G, T6
Class I, Zone 2 AEx nA II, T6
CSA
Ex nA II T6
Ex nA IIC T6 Gc
NEPSI
(China)
Intrinsically Safe & Non-Incendive
Custom Union including Russia
"On safety of the equipment operating
under excessive pressure"
Custom Union including Russia
"On safety of the equipment for work in
explosive environments"
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
All manuals and user guides at all-guides.com
Limit Switches - Inductive Alarm Switch
One or two electronic limit switches type SJ2-N can be installed in the
indicator housing to allow initiation of signaling or switching functions on a
preset flow value being reached. The SJ2-N limit switch operates as a slot
initiator that is inductively actuated by a cam mounted to the pointer. Any
flow value can be used for setting the limit value by sliding the switch along
the slot in the mounting plate for the initiators. Minimum setting distance
between two limit switches is approximately 50% of the scale range.
Power supply
Current consumption
Current consumption
Self inductance
Self capacitance
Max Temp
The flow valve can be used for setting the limit value by sliding the switch
along the slot in the mounting plate for the initiators. Minimum setting
distance between two limit switches is approximately 50% of the scale
range.
Applicable Standard
EN 60079-0:2006
EN 60079-11:2007
EN 61241-0:2006
EN 61241-11:2006
IEC 60079-0:2007-10
IEC 60079-11:2006
75°C
Db IP65
IEC 61241-11:2005
UL & CSA Standards
GB3836.1-2010
GB3836.4-2010
UL & CSA Standards
GB3836.1:2010;
GB3836.8:2003
TR CU 032/2013
R U 012/2011
(TR CU Ex)
Section 1 Introduction
Brooks
Ar-Mite
®
8 Vdc (Max. 15.5 Vdc)
active area clear: > 3 mA
active area obscured: < 1 mA
29 μH
20 nF
158°F (70°C)
Details
KEMA 02ATEX1126
IECEx KEM 09.0046
1379260
GYJ11.1639
1379260
GYJ13.1315
TC N RU -
U.A 04.B.05988
RU C-HU.
08.B.00741
MT3750C
TM
1-7

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

541b063aag

Tabla de contenido