B. Installation and setup

Appendix B. Installation and setup

B1. System installation

B1.1 Positioning the unit

B1.2 Positioning the PC(s) (not supplied)

Installation and setup must only be carried out
by service personnel trained and approved by
the manufacturer of the unit.
Position the unit on a stable fl at surface so that
brations will not degrade the image quality.
The unit can also be attached to a wall, under or on
a shelf using the optional scanner mounting kit
(pt. no. 204138).
Do not position the unit in direct sunlight or near
bright light. Sunlight or bright light must not be al-
lowed to shine directly on the unit door into which
the IPs are inserted.
The unit must not be positioned so that it touching
other equipment. The unit must not be placed on top
of or under other equipment.
The unit can be positioned within the environment in
which the patient is examined and treated (patient
environment).
The PC(s) connected to the unit should not be used
in the patient environment.
The minimum horizontal distance between the pa-
tient and the PC(s) is 1.5 m (4.5 ft).
The minimum vertical distance between the patient
and the PC(s) is 2.5 m (6.5 ft).
Imaging plate unit
vi-
B - 1