KaVo eXam 0.805.0333 Manual Del Usuario página 53

Sistema de radiología intraoral digital
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

B. Installation and setup
Obtain an IP address for the unit from your net-
work administrator and key it into the IP fi eld in
the Scanner IP address area.
NOTE! The PC and the unit must be in the
same subnet when setting the IP address of the
unit.
3. PC + Unit: Press and hold down the Start key
on the unit and then click the Send to Scanner
button on the settings window.
You will hear a beep which indicates that the
PC is now sending the IP address the unit. The
Startup animation will appear on the unit display
4. PC: When the Startup animation clears click
OK to connect the PC to the unit.
5. Now connect the other PCs in the network to the
unit. Just enter the IP address into the IP fi eld
and then click OK to connect the PC to the unit
(it is not necessary to hold down the Start key
and click the Send to Scanner button with the
other PCs once the unit has already got an IP
address).
Imaging plate unit
B - 5

Publicidad

Tabla de contenido
loading

Tabla de contenido