MEMORY FUNCTIONS
All manuals and user guides at all-guides.com
After measurement, the last set of measured values will be saved automatically, when the
unit is turned off by pressing "POWER" button or by automatic power-off after 3 minutes.
View past readings
1. Press the "memo s" button to enter memory function.
2. If the User Page is not what you want to view, press the "paget" button to select
average function.
3. Press the "memo s" button to display the next set of values. Past reading are displayed
from the most recent to the oldest.
4. This can store up to 60 past readings for each user (2 x 60)
TROUBLESHOOTING
Problem
No display when you press
the Power/Start button
"Err P" displayed
"Err 1" displayed
"Err 2" displayed
"Err 3" displayed
Note: If your problem cannot be solved by the above, consult your store of purchase.
Do not disassemble the unit.
STORAGE & MAINTENANCE
• Keep away the unit from direct sunlight, extreme temperatures, humidity or moisture.
• Use a dry, soft cloth to clean the unit, or if desired, use a cloth lightly dampened with water.
• Do not use alcohol, benzene, thinner or other volatile liquids to clean the unit.
• Do not wash or expose the wrist cuff to liquid.
• Remove batteries from the unit when not in use for more than 3 months.
SPECIFICATIONS
Measurement method
Measurement range
Accuracy
Inflation
Pressure detection
Power supply
Auto power off
Storage environment
Operation environment
External dimensions
Net weight
Cuff dimensions
Cause
Are the batteries done?
Are the battery polarities wrong?
Fails to inflate
Deflates too rapidly
Movement during measurement
Signal interference
Oscillometric
Pressure: 20 – 280 mmHg
Pulse Rate: 40 – 195 beats/min
Pressure: +/- 3 mmHg
Pulse Rate: +/- 5%
Pump driven
Semi-conductor
4 x AA batteries
3 minutes after last operation
-20°C to 60°C, 10% to 95% RH
10°C to 40°C, 30% to 85% RH
195 (L) x 165(W) x 87 (H) mm
602g (batteries excluded)
Arm circumference of 22~34cm
Solution
Replace with new batteries.
Re-align batteries with correct polarities.
Remove interfering source e.g. mobile
phones, magnets. Measure again.
Return for servicing.
Do not move during measurement
Check if the cuff is properly connected.
Replace the air tube if it is broken.
Incorrect results