View average readings
1. Press the "
AVG" button to enter average function.
2. If the User Page is not what you want to view, press the "PAGEt" button to select
average function.
3. Press the " AVG" button to view different average values. 5 different average values are
available:
Delete past readings
The user can delete all the past readings stored in each User page.
1. Press the "PAGEt" button to select the User page you want to delete.
2. Press and hold the "sMEMO" & "PAGEt" buttons simultaneously for 5 seconds. The
symbol of "ALL EE" will be displayed and all past readings stored in that User Page will be deleted.
TROUBLESHOOTING
Problem
No display when you press
the Power/Start button
"Err P" displayed
"Err 1" displayed
"Err 2" displayed
"Err 3" displayed
Note: If your problem cannot be solved by the above, consult your store of purchase. Do not
disassemble the unit.
All manuals and user guides at all-guides.com
Symbol on Display
1h
1d
7d
30d
ALL
Cause
Are the batteries done?
Are the battery polarities wrong?
Fails to inflate
Deflates too rapidly
Movement during measurement
Signal interference
Average Values for
One hour
One day
One week
One month
All readings
Solution
Replace with new batteries.
Re-align batteries with correct polarities.
Remove interfering source e.g. mobile
phones, magnets. Measure again.
Return for servicing.
Do not move during measurement
Check if the cuff is properly connected.
Replace the air tube if it is broken.
Incorrect results