Emergency opening of centrifuge lid

To open the lid of centrifuge in case of power break-
down or lid lock damage, disconnect the device from
mains power supply, wait until the rotor comes to a
standstill then lift the centrifuge and insert the in-

Rotor assembly and usage

Assembly: Place the rotor on the axis . On top of the
rotor place a special spacer - note that the pin of the
spacer must align with the groove of the axis - then
place the nut on the axis and tighten it with the in-
cluded rotor key . Before each start, check that the
rotor is firmly tightened.
Disassembly is performed in reverse sequence .
Usage: Always load the rotor symmetrically . By
minimizing the difference of weight between the
filled test tubes, you reduce the wear of mechanical
parts of the centrifuge . If the overall imbalance of
Centrifuge CM-8S
USER MANUAL
cluded emergency lid opening key in to the groove
and move the lever to the side (as it is shown on the
figure below). The lever is located on the bottom of
centrifuge in the groove near the control panel .
EMERGENCY LID OPENING LEVER
test tubes in the rotor exceeds 7 grams, centrifuge
will start emergency braking it reaching 1000 rpm
and the display will show when an imbalance error .
ATTENTION! Do not use damaged rotors!
ATTENTION! Samples may contain pathological mate-
rial, including pathogens of serious diseases. Always
check test tubes for damage before centrifugation.
7
ENG