Error code table
Error code
Cause
No main power connection.
Power failure
Centrifuge lid is left open.
Total imbalance of test tubes is more
than 7 gram.
Processor Control Board is overheated.
Residual rotation of the rotor.
Centrifuge CM-8S
USER MANUAL
Solution
Check connection of the power cable.
Close and press on the lid till you hear a
click.
Load rotor symmetrically.
Give a little time and centrifuge will cool
down automatically.
Wait until the rotor is motionless.
9
ENG