Data Management; Menu Structure Of Data Menu; Measurement Data; Calibration Data - Mettler Toledo SevenGo Duo pro SG68 Instrucciones De Manejo

Ocultar thumbs Ver también para SevenGo Duo pro SG68:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7 Data management

7.1 Menu structure of data menu

1.

Measurement Data

1. Review
2. Delete
2.

Calibration Data

1. pH
1. Review
2. Delete
2. Ion
1. Review
2. Delete
3. DO
1. Review
2. Delete
7.2 Measurement data
Review
All
All stored measurement data can be reviewed; the most recent data saved appears on the display.
Partial
The measurement data can be filtered according to 3 criteria.
Memory number (MXXX)
Sample ID
Measurement mode
Memory number
Enter the memory number of the data and press Review.
The measurement data is displayed.
Sample ID
1 Enter the sample ID and press Review.
The meter finds all stored measurements with this sample ID.
2 Scroll through the measurement data to review all measurements with the entered sample ID.
Measurement mode
1 Select a measurement mode from list and press Review. The meter finds all stored measurements of the
selected measurement mode.
2 Scroll through the measurement data of the selected measurement mode.
Delete
All or partially stored measurement data can be deleted by filtering the measurement data. The filter works as
described above in "Review".
Note
Deletion is protected by a PIN. Upon delivery, the PIN is set to 000000. Change the PIN code to prevent
unauthorized access.
7.3 Calibration data
Calibration data can be reviewed and deleted. Up to 5 calibrations per sensor ID are stored in the memory.
24
Data management​ ​
3.

ISM Data

1. pH
1. Initial Calibration Data
2. Calibration History
3. Maximum Temperature
4. Reset ISM
2. DO
1. Initial Calibration Data
2. Calibration History
3. Maximum Temperature
4. Oxygen Membrane
5. Reset ISM
SevenGo Duo pro™

Publicidad

Tabla de contenido
loading

Tabla de contenido