1 Select between the sensor types: pH, ion or DO.
2 Press Review.
A list of calibrated sensor IDs appears.
3 Select a sensor ID from the list and press Review.
4 Press
and
— or —
Press and hold CAL for 3 seconds in the single channel measurement screen.
The current calibration data is displayed.
Delete
1 Select between the sensor types: pH, ion or DO.
2 Press Delete.
A list of sensor IDs appears.
3 Select a sensor ID from the list and press Delete.
4 Press Yes when the message "This will delete all your selected data. Please confirm." appears
— or —
Press No to cancel and exit.
After deletion, the sensor ID disappears from the list in the sensor ID menu.
Note
An active sensor ID cannot be deleted.
This menu is protected by a deletion PIN code. Upon delivery, the PIN code is set to 000000. Change the
PIN code to prevent unauthorized access.
7.4 ISM data
SevenGo Duo pro™ meter incorporates Intelligent Sensor Management (ISM
functionality provides extra security, safety and eliminates mistakes. The most important features are:
Extra security!
After connecting the ISM
number are transferred from the sensor chip to the meter.
After calibration of the ISM
chip. The most recent data is always stored where it should be – on the sensor chip!
Extra safety!
After connecting the ISM
reviewed to see the development of the sensor over time. This information provides an indication if the sensor
should be cleaned or renewed.
Eliminate mistakes!
After connecting an ISM
Additional features are described below.
Initial calibration data
®
When an ISM
sensor is connected, the initial calibration data in the sensor can be reviewed or transferred. The
following data is included:
Response time
Temperature tolerance
Membrane resistance
Slope (at pH 7) and offset
Type (and name) of electrode (for example, InLab Expert Pro ISM
Serial number (SN) and ordering (ME) number
Production date
SevenGo Duo pro™
to navigate between the previous or next calibration data sets.
®
sensor, the sensor is automatically recognized and the sensor ID and serial
®
sensor, the calibration data is automatically stored from the meter to the sensor
®
sensor, the five most recent calibrations are transferred to the meter. These can be
®
sensor, the last set of calibration data is automatically used for measurements.
®
) technology. This ingenious
®
)
Data management​ ​ 25