Final X to Y Check

As a final check, compare X and Y axes to be sure that your adjusted
calibration is within the specs of ± 3/32". The marks for X, X', Y, and Y'
should be no more than 3/16" apart at 100 ft. (5 mm at 30 m).
wc_tx000805gb.fm
Select the Y axis by pressing the HI Alert Function key (10)
on the laser keypad (on the remote, press the double arrow
key >>l ).
Y LED will blink rapidly, indicating leveling. When the LED
blinks slowly, the laser is ready to calibrate.
If you have not moved the laser, use the Y marks made in
previous steps of "Checking Y axis".
Use arrows on the keypad to move the beam up or down to
the halfway mark. If the Y axis is toward the wall with the
marks, use the ^ key (15) to raise the beam, and the v key
(16) to lower the beam.
Press the MAN key to save the calibration you've just made
on the Y axis. (On remote, use key with the small dot).
EN-24
Maintenance