iQC
TM
The analyzer features an internal quality control (iQC
suppresses values and displays warning or error messages when a problem
occurs. The iQC
TM
every run while monitoring all technical components of the analyzer.
A suppression or warning does not necessarily mean that your instrument
has a technical fault. It proves that your iQC
ensures that only accurate results will be released.
The symbols ~~~ are printed in place of numbers when a result
•
cannot be determined. The result may be suppressed due to improper
mixing of a reagent bead with diluted sample, a nonlinear reaction, an
endpoint of a particular reaction not reached, a concentration outside
the analyzer's capabilities or thermal stress. When a chemistry is
suppressed (~~~), the analyzer prints an error report.
HEM, LIP, or ICT is printed in place of the analyte concentration if
•
hemolysis, lipemia or icterus have adversely affected the results.
LIP is also printed if both lipemia and icterus have affected a particular
analyte. HEM is also printed if hemolysis and icterus, hemolysis and
lipemia, or hemolysis, lipemia, and icterus have affected a particular
analyte. Examine the sample indices to determine if more than one
interferent is affecting a particular result.
CONFIRM LOW RECOVERIES with an exclamation point (!) next to
•
every analyte indicates that at least one analyte has a lower concen-
tration than normally expected. If this occurs, re-run the sample —
if the message persists, the sample may be problematic.
Warnings and error messages include an internal error code that will
•
assist the Abaxis Technical Support in diagnosing the issue. Record
the error message and/or print the error report (see chapter 5.2 of the
operator's manual) before contacting Abaxis Technical Support:
9