Measuring AC (RMS) or DC current up to 100 mA
– Measurement Range (AC): 10.0000 mA, 100.000 mA
– Measurement Range (DC): 100.000 μA, 1.00000 mA, 10.0000 mA, 100.000 mA
– Speed (AC): Slow-2 Hz, Medium-20 Hz, Fast-200 Hz
– Speed (DC): Slow, Medium, Fast
– Default Setting: Autoranging, Slow measurement speed
– Shunt Resistance: 1 for 10 mA and 100 mA, and 90 for 100 μA to 1 mA
– Input Protection: Rear Panel 0.4 A, 500 V FH fuse for the I terminal