Functions and Features
Trigger Out
The Trigger out signal cannot be configured and is implemented in four ways:
– In local mode, a trigger out signal is sent whenever a measurement is updated
– In remote mode, a trigger out signal is sent whenever a measurement is taken
– In data log/external trigger mode, a trigger out signal is sent whenever a
– In code compatibility mode, a trigger out signal is sent whenever a
Figure 2-6
Manual Trigger
Manual triggering takes one reading (or the specified number of readings in data
logger) each time the multimeter receives a pulse on the front-panel trigger
button. The diagram below shows the manual trigger operation:
82
on the front panel.
by the user through using command.
measurement is log/triggered and updated on the front panel.
measurement is updated on the front panel or a measurement is taken by the
user using command.The diagram below shows the trigger out connector:
Trigger out connector
TRIG
TRIG
Tx
Rx
IN
OUT
–
+