4.
Product description
4.1
Scope of supply
Art. no.
Designation
GA336
Oscillating saw straight
– or –
GA337
Oscillating Sternum saw
GA678
Sterile funnel
TA014547
Instructions for use of oscillating saw straight GA336 /
Oscillating Sternum saw GA337 (flyer)
4.2
Components required for operation
■
Battery GA676 (charged)
■
Sterile funnel GA678
■
Cap GA675
■
Saw blade (according to indication)
4.3
Operating principle
The product 1 contains an electric motor, which is powered by a replace-
able battery 7.
The charged non-sterile battery 7 is inserted into the product 1 via the
sterilized funnel 6 and sealed with the cap 8 so that it is sterile.
The oscillation frequency is controlled electronically and can be continu-
ously regulated with the trigger 2.
The product 1 has a coupling, which allows for the attachment of various
saw blades 11. These saw blades 11 lock automatically when attached to
the product.
The saw blade 11 can be removed by activating a push button 13.
5.
Preparation
Aesculap assumes no liability if the following rules are not followed:
►
Do not use products from open or damaged sterile packaging.
►
Prior to use, inspect the product and its accessories for any visible
damage.
►
Use the products and their accessories only if they are in perfect tech-
nical condition.
6.
Working with the device
Risk of infection and contamination!
Product is delivered unsterilized!
WARNING
►
Sterilize the product before use pursuant to the
operating instructions.
Risk of injury and material damage due to acciden-
tal activation of the product!
►
Products which are not being actively used must
WARNING
be secured against accidental activation (posi-
tion OFF).
Risk of injury and material damage due to inappro-
priate use of tools!
►
WARNING
Always follow the safety advice and information
given in the instructions for use.
►
When coupling/uncoupling, handle tools with
cutting edges with care.
Damage to the product if dropped!
►
Use the products only if they are in perfect
technical condition, see Function check
WARNING
Risk of burns to skin and tissue through blunt tools
or if product has not been maintained properly!
►
Use tools only if they are in perfect condition.
WARNING
►
Replace blunt tools.
►
Maintain the product properly, see maintenance
guide.
5