Oscillating saw straight GA336 / Oscillating Sternum saw GA337
6.1

System set-up

6.1.1

Connecting the accessories

Risk of injury due to unapproved configuration
using additional components!
DANGER
Ensure that the classification of all components
used is consistent with the classification of the
product (such as type BF or type CF).
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
All
configurations
must
IEC/DIN EN 60601-1. The person connecting the devices with each other
is responsible for the configuration and must ensure that the fundamental
standard IEC/DIN EN 60601-1 or relevant national standards.
Follow the instructions for use of individual accessories.
Please address your B. Braun/Aesculap partner or Aesculap Technical
Service with any inquiries in this respect; for a contact address, see
Technical Service.
6.1.2

Inserting the rechargeable battery

The product does not function when using the
wrong battery!
CAUTION
Do not use battery GA666 (with red base) or
GA346.
Product 1 with the battery shaft 4 must be turned upwards and the
sterile funnel 6 (sterile) attached, see Fig. A.
Battery 7 (non-sterile) is inserted into the battery shaft 4 by a second
(non-sterile) person, see Fig. A.
Note
After the battery has been inserted, multiple signals will sound, which indi-
cate that the product is ready to use.
After insertion of the battery, have the Sterile funnel 6 (non-sterile)
removed by a second person.
Attach the cap 8 (sterile) in such a manner that it clicks into place with
both release catches 9.
Note
The sterility of the product is only guaranteed with a correctly placed cap.
6
Acculan 4
®
fulfill
the
fundamental
6.1.3

Intraoperative battery change

WARNING
The battery removal device is used for intraoperative battery changes
ensuring that sterile conditions are safely maintained.
Turn the product 1 with the battery shaft 4 upwards.
Press both safety catches 9 on the cap 8 at the same time and remove
the cap 8.
Attach the sterile battery removal device 10, see Fig. B.
Shake the product 1 with the attached battery removal device 10 with
standard
battery shaft 4 slightly downwards.
The battery 7 will slide easily into the battery removal device 10.
Give the battery removal device 10 including dead battery 7 to the
non-sterile person.
Insert a charged battery 7, see Inserting the rechargeable battery.
6.1.4

Removing the rechargeable battery

WARNING
CAUTION
CAUTION
Rechargeable battery must be removed after each surgical procedure and
prior to processing of the device.
Note
The battery removal device can be used for easier battery removal, see
Fig. B.
Risk of injury from hot battery!
Battery may be hot after use in the machine.
Remove the battery with the battery removal
device and let it cool.
Risk of injury from hot battery!
Battery may be hot after use in the machine.
Let the battery cool in the machine and then
remove.
– or –
Remove the battery with the battery removal
device and let it cool.
Damage to the battery caused by knocking it
against hard objects!
Only remove battery by tapping the product on
the palm of the hand.
Damage to, or destruction of the batteries due to
processing!
Do not sterilize batteries.