BAR model
Removing the 350g weight and putting a 300g weight on the scale the display will show 85.71% as 300g is 85.71% of
350g.
Note: the scale may jump by large numbers unexpectedly if small weights are used to set the 100% level. For example
if only 23.5g is on a scale with 0.5g increments and the scale is set to 100%, the display will show 100.00%, however a
small change of weight will cause the display to jump to 102.13% as one scale division (0.5g) increase to 24.0g will be
equivalent to a 2.13% increase.
Removing the 350g weight and putting a 300g weight on the scale the display will show 85.71% as 300g is 85.71% of
350g.
Pressing the
key will return the scale to weighing.
5.5. PARTS COUNTING
When the scale is showing weight, pressing the
Before beginning, tare the weight of any container that will be used, leaving the empty container on the scale. Place
the number of samples on the scale. The number should match the options for parts counting, 10, 20, 50, 100 or 200
pieces.
Press the SMPL key to begin. The scale will show "SP 10" asking for a sample size of 10 parts. You can press
select sample quantity: 10, 20, 50, 100, 200 and back to 10.
Press the
when the number matches the number of parts used for the sample. As more weight is added the display
will show the number of parts (pcs).
Press the
key to display unit weight (g/pcs), Total weight (kg) or the count (pcs). Press the
normal weighing.
5.6. ACCUMULATION
When the scale is set to manual accumulation the weight displayed will be stored in memory when the
and the weight is stable.
The display will show "ACC 1" and then the total in memory for 2 seconds before returning to normal. If the optional
RS232 interface is installed the weight will be output to a printer or PC.
Remove the weight, allowing the scale to return to zero and put a second weight on. Press
"ACC 2" and then show the new total. Continue until all weights have been added.
To view the totals in memory press the
"ACC xx" and the total weight before returning to zero. The totals will also be printed via the RS232 interface.
To erase the memory press
* When you don't connect to printer, please don't set RS232 to LP50 mode, otherwise accumulate operate will be slow.
See detail in section 8.
6. BATTERY FONCTION
The balance come with a rechargeable battery, can be operated from the battery if desired. The battery life is approxi-
mately 40 hours.
There is a battery symbol on display used to indicator battery status:
Flick: charging battery
Full: battery full or charge finished
Half: battery with half power
Blank: need be charged
key will start the parts counting function.
key when the scale is at zero. The display will show the total number of items
to view the totals and then press the
, the display will show
key to clear the memory.
User's Guide
key to
key to return to
key is pressed
25