WARNING: Risk of Electric Shock - Connect this HCOA to a properly grounded Hospital Grade ("green dot")
Outlet ONLY"
• Do not use outdoors.
• Do not place on floors.
WARNING: Risk of electrical shock. This product is only intended for use with a properly grounded duplex
receptacle outlet. Do not use in a single receptacle outlet or remove the grounding pin.
DANGER: Risk of Death or Serious Injury. Connect only AUTHORIZED medical equipment that has
been verified for safe touch and leakage current.
WARNING: After verifying the combined touch and leakage current, secure the unused receptacle covers
closed.
• The HCOA shall be used within its marked load rating.
• Use of this equipment in life support applications where failure of this equipment can reasonably be
expected to cause the failure of the life support equipment or to significantly affect its safety or
effectiveness is not recommended. Do not use this equipment within oxygen-enriched atmospheres, or
within 0.3 m of a point at which an oxygen-enriched atmosphere is intentionally vented.

Grounding Instructions

This product must be grounded at all times when in use and when used within a Patient Care Vicinity. If malfunction or
insulation breakdown occurs, grounding provides a path of least resistance for electric current to reduce the risk of
electric shock. This product is equipped with a cord that contains a special integral equipment-grounding conductor and
grounding plug. The plug must be connected to a duplex outlet that is properly installed and grounded in accordance with
all local codes and ordinances. This product also includes an attachment plug with integral patient equipment grounding
pin, which must be connected to the duplex receptacle before use within any Patient Care Vicinity (6 ft. / 1.8 m), beyond
the patient bed, chair, table, treadmill or other device that supports the patient during examination and treatment.
DANGER - Improper connection of the equipment-grounding conductor and Patient Equipment Grounding Conductor can
result in a risk of electric shock. Check with a qualified electrician, technician or service person if you are in doubt as to
whether the product is properly grounded.
Mounting
This HCOA is intended to be mounted to benches, stands, carts or other areas containing multiple cord-and-plug
connected medical electrical equipment that is intended and authorized for use in a patient care (both patient care
"category 2" or critical care "category 1") location.
Mounting may be permanent (requiring a tool to remove) or temporary (can be removed from without the use of a tool,
i.e., keyhole slots).
Mount and position to prevent any liquid from entering the HCOA.
Tripp Lite has a policy of continuous improvement. Specifications are subject to change without notice. Photos and illustrations may
differ slightly from actual products.
1111 W. 35th Street, Chicago, IL 60609 USA • tripplite.com/support
2
21-08-259 93-4064_RevA