4.Operation Instruction
Danger
In measurement of earth voltage, do not exert voltage above AC220V on the measurement port.
In measurement of earth resistance, there will occur potential of about 50 V between port E and S or between ports E and C,
therefore, protect yourself from electrical shock.
Before using the meter, check the electricity in the battery by turning on the meter, checking if there is "
battery by following the steps described in Chapter "Replacement of the battery" if the symbol exists.
4.1 Measurement of earth voltage
Rotate the switch to ~VE position, insert the end of meter cable into sockets E and S, with another end of meter cable connected to either
end of voltage source or resistance load in parallel way to measure . If there exists voltage, it will be displayed on LCD.
4.2 How to measure earth resistance by Bi-pole method
Existing earth pole, such as buried metals (metal water pipe etc) , common earth of the commercial power supply or lightning rods of high
buildings, instead of auxiliary earth bars are used as auxiliary pole in this method.
8
" on display and replace the