Descargar Imprimir esta página

8- Safety Devices - ACTEON Equipment SATELEC Implant Center Manual De Uso

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 52
c) Reboot in factory configuration
Confirm access to setting
a check mark is displayed next to the selected
setting.
Confirm the return to the factory configuration
by pressing button
d) Software version
The version of the embedded software (X.Y) is
displayed next to the symbol
The I
C
is configured in the factory
MPLANT
ENTER
with the following settings:
Motor mode:
Contra-
Program
Function
angle ratio
1
Marking
20 :1
Pilot
2
20 :1
drilling
3
Tapping
20 :1
4
Screwing
20 :1
P
mode:
IEZOTOME
Program
Function
1
D1
2
D2
3
D3
4
D4
N
mode:
EWTRON
Program
Function
1
Green
2
Yellow
3
Blue
4
Orange
All manuals and user guides at all-guides.com
by pressing button ;
.
.
Speed at
Torque at
instrument
instrument
Irrigation
end
end
1200
80 N.cm 80 ml/min
tr/min
800 tr/min 80 N.cm
ml/min
15 tr/min 40 N.cm
ml/min
30 tr/min 40 N.cm
Modulation
Irrigation
frequency
60 Hz
50 ml/min
30 Hz
50 ml/min
10 Hz
50 ml/min
30 Hz
50 ml/min
Power
Irrigation
Soft
P = 5
15 ml/min
Medium
P = 5
15 ml/min
High
P = 5
15 ml/min
Boost
P = 5
15 ml/min
VIII - SAFETY DEVICES
The I
C
MPLANT
malfunctions of the device.
When an excessively high temperature of the I-
S
micro-motor is detected, symbol 1 in Fig. 9
URGE
is displayed in the error zone (Fig. 4-13) and the
device beeps 3 times.
The device switches to degraded mode so that it
can complete the dental procedure.
The available torque value is set at 25% in order
to protect the I-S
It is recommended that the I-S
be allowed to cool down until symbol 1 in Fig. 9
disappears.
100
100
When an internal operation error occurs, symbol
2 in Fig. 9 is displayed in the error zone and the
device beeps 3 times.
0
It is recommended that the device be switched
off (Fig. 2-5) and switched back on again.
has a system that detects
ENTER
micro-motor.
URGE
URGE
micro-motor
15