TROUBLESHOOTING
PROBLEM
The Orä fails
to power on.
"ER1" is
displayed.
The temperature
reading is
lower than the
typical body
temperature
range.
The temperature
reading is
higher than the
typical body
temperature
range.
14
POSSIBLE CAUSE
Low battery
The polarities of the
batteries are reversed.
The themometer is
damaged.
The ambient temperature is
lower than 50 °F (10 °C) or
higher than 104 °F (40 °C).
The lens of the temperature
probe is dirty.
The thermometer probe is
not aligned to the eardrum.
The thermometer is used
within 30 minutes after
being taken from a cold
environment.
The temperature probe
is damaged.
SOLUTION
Change the batteries
Make sure that
the batteries are
installed correctly.
Contact the
manufacturer.
Take the measurement
under an ambient
temperature between
50 °F (10 °C) and
104 °F (40 °C).
Clean the lens using
a cotton swab.
Reposition the thermometer
probe so that it is aligned to
the eardrum.
Wait for more than
30 minutes after the
thermometer is moved
into the measurement
environment.
Contact the
manufacturer.