To abort calibration, press calibration button. It is forbidden to 
abort calibration manually when "CAL" is displayed on the screen 
 
 
INSTRUCTIONS FOR USE 
mean that automatic calibration is currently performed. Wait 
about 15s and try to press a button again. 
measurement sensor. The sensor must be connected to the 
socket on the device, and then device RESTART must be 
performed (switch off and on again). Symbol 
the screen. 
flashes or "oFF" appears on the screen, batteries should be 
replaced with new ones.  
the table provided in the further part of this instruction manual. 
 
 
 
 
 
3.4.
Cleaning and maintenance 
 
a)
Before cleaning and when the device is not in use, pull the power supply plug. 
b)
Use only non‐corrosive agents for cleaning the surface. 
c)
Store the device in a dry and cool place protected from moisture and direct 
sunlight. 
d)
Do not spray the device with a stream of water and do not immerse it in water. 
e)
Ensure that no water enters through the ventilation openings in the housing. 
400ppm level calibration 
‐ Set the sampling frequency to 5s 
‐ Unscrew the screw and remove the flap in the back of 
the housing to gain access to the calibration button. 
‐ Press and hold the calibration button, until "CAL 400" 
appears on the screen. 
‐ Set the meter in optimum environmental conditions. 
‐ After 5 min the device will perform calibration 
automatically, and when the process is finished it 
automatically returns to the main screen. 
When the buttons of the device do not respond it may 
The device is adapted for connection of an external 
In the even when the battery symbol on the screen 
In the event error code appears on the screen, refer to 
 will appear on 