as replacement parts for internal components, may result in increased
emission or decreased immunity of the device.
3.WARNING:The medical devices should not be used adjacent to
or stacked with other equipment. In case adjacent or stacked use is
necessary, the medical device should be observed to verify normal
operation in the configuration in which it will be used.
4.PORTABLE RF communications equipment(including peripheras
such as antenna cables and external antennas)should be used no
closer than 30cm(12 inches)to any part of [ME QUIPMENT or ME
SYSTEM], including cables specified by the MANUFACTURER.
Otherwise,degradation of the performance od this equipment could
result.
5.Do not use mobile (cellular) telephones and other devices(such
as MRI, diathermy, electrocautery, RFID and electromagnetic security
systems) which generate strong electrical or electromagnetic fields, near
the medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to keep a
minimum distance of 7m. Verify correct operation of the device in case
the distance is shorter.
Cables
USB Cable
Guidance and Manufacturer's declaration – electromagnetic emissions
This device is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should assure
that it is used in such environment.
Electromagnetic emission IEC 60601-1-2
Emissions test
RF emissions
CISPR 11
Length ( m )
1
Compliance Electromagnetic environment – guidance
Group 1
24
Whether to block
No
This device uses RF energy only for
its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
Note
-