RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and Manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should assure
that it is used in such an environment.
IEC 60601 test
Immunity test
level
Electrostatic
discharge
±6 kV contact
(ESD) IEC
±8kV air
61000-4-2
±2 kV for power
Electrical fast
supply lines
transient/burst
±1 kV for
IEC 61000-4-4
input/output lines
±1 kV line to
Surge IEC
line
61000-
±2 kV line to
4-5
earth
Class B
This device is suitable for use
in all establishments, including
domestic establishments and those
Class A
directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
Complies
purposes.
Compliance
level
±6 kV contact
±8kV air
±2 kV for
power
supply
lines*1)
±1 kV line to
line
±2 kV line to
earth
25
Electromagnetic
environment – guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.