Mounting and Removing the Adaptor for Oscillating Intra Oral Blade (Optional)
_ Mounting
1
While pulling down the Removal Lever Ⓐ, fit the concave part of
the Adaptor for Oscillating Intra Oral Blade Ⓓ into the convex
part of the Attachment Ⓒ.
2
Release the Removal Lever Ⓐ to set the Adaptor for Oscillating
Intra Oral Blade in place.
_ Removing
Perform steps 1 and 2 of the mounting procedure in the reverse order.
Mounting and Removing the Cutting Accessories (Optional)
1
Ensure the Screw part of the Adaptor for Oscillating Intra Oral
Blade Ⓕ and Notch Section on the Cutting Accessory Ⓔ are
located (12 o-clock) upward. Then insert the Cutting Accessory
into the Adaptor for Oscillating Intra Oral Blade.
2
Rotate the Special Wrench Ⓖ provided with the Adaptor in the
side hole of the Adaptor to set the Cutting Accessory Ⓗ in place.
_ Removing
Perform steps 1 and 2 of the mounting procedure in the reverse order.
3
By twisting the Removal Lever Ⓐ clockwise and
counterclockwise, set the Cutting Accessory in place properly.
* If the Cutting Accessory is not set in place properly, repeat the
procedure from 1 - 3 to set it in place properly.
3
After mounting the Cutting Accessory Ⓗ, pull it lightly to confirm
that it is securely connected to the Attachment.
* If the Cutting Accessory is not set in place properly, repeat the
procedure from 1 - 3 to set it in place properly.