Regulatory Information
Important Notice for Prospective Sound Generator Users
Good health practice requires that a person with tinnitus have a medical evaluation
by a licensed physician (preferably a physician who specializes in diseases of the ear)
before using a sound generator. Licensed physicians who specialize in diseases of the
ear are often referred to as otolaryngologists, otologists or otorhinolaryngologists.
The purpose of a medical evaluation is to assure that all medically treatable
conditions that may affect tinnitus are identified and treated before the sound
generator instrument is used.
44
Regulatory Information
TINNITUS TECHNICAL DATA
Custom Tinnitus
Technology Maximum Output = 87 dB SPL (typical) when measured
in a 2cc coupler per ANSI S3.22 or IEC 60118-7.

EMC COMPLIANCE

This hearing aid model has been tested to, and has passed, the following emissions
and immunity tests:
• IEC 60601-1-2 radiated emissions requirements for a Group 1 Class B device as
stated in CISPR 11.
• RF radiated immunity at a field level of 10 V/m between 80 MHz and 2.7 GHz as
well as higher field levels from communications devices as stated in Table 9 of
IEC 60601-1-2.
• Immunity to power frequency magnetic fields at a field level of 30 A/m and
proximity magnetic fields as stated in Table 11 of IEC 60601-1-2.
• Immunity to ESD levels of +/- 8 kV conducted discharge and +/- 15 kV air
discharge.