HCB-200-425F Specification
The figures stated in the specification are as measured 0-50 mm from the base of the well of the insert,
using an SPRT, at the time of calibration.
Minimum temperature
Maximum temperature
Temperature accuracy
Temperature uniformity
Temperature stability at 200 after 10 minutes
Temperature stability at 425 after 10 minutes
Display resolution
Set point display resolution
Heat up 12 minutes
Cool down 21 minutes
Fan cooling Automatic
Dimensions Height x Width x Depth
Large well in the block:
Weight
Options
Central Well for inserts, see the list of inserts
Full bi-directional RS232 Comms Port 9way D type plus software and cable
Soft carry case
Hard carry case
UKAS calibration certificate
20°C above ambient
425°C
±0.3°C
±0.2°C
±0.03°C
±0.05°C
0.01°C
0.1°C
20°C to 400°C
400°C to 100°C
Automatic
270mm x 170mm x 255mm
Depth / Diameter 130mm / 38.2mm
6.3Kg
15

Publicidad

loading

Este manual también es adecuado para:

Hbc-200 serie