Descargar Imprimir esta página

Uline H-9884 Manual página 4

Publicidad

Idiomas disponibles

Idiomas disponibles

USER MENU
NOTE: The user menu allows the customizing of
scale settings.
NOTE: All modes except for weighing must be
activated in the user menu before they are
available.
Sub -
.C.a.l.
.S.e.t.u.p.* .M.o.d.e
Menus
Menu
Span
b.light
Items
LIN
A-OFF
Filter
AZT
Stab
Stab.C
End
End
1.
Press and hold MENU until MENU is displayed. When
released, the first sub-menu (.C.A.L.) will be shown.
2. Press YES to enter the displayed sub menu or press
NO to advance to next.
3. Press YES to view the menu item setting or press NO
to move to the next menu item. When viewing the
setting, press YES to accept the setting or NO to
change the setting.
4. When ENd is displayed, press YES to return to the
sub-menu selections or NO to return to the first item
in the current menu.
CALIBRATION SUB-MENU
• Span [SPAN] (yes, no) – Initiates a span calibration
procedure (zero and span). A span calibration is
important when initially setting up the scale.
• Lin [LIN] (yes, no) – Initiates a linearity calibration
procedure (zero, mid-point and span).
PAGE 4 OF 24
.U.n.i.t.*
.E.n.d
Count
g, kg, N
Check
oz, ozt, dwt
lb, lb:oz, ct
grain
End
End
SETTINGS
SETUP SUB-MENU
• Back Light [b.L ight] (on, off, Auto) – When Back light
is set to "ON", the scale will always be on.
• Back Light [b.L ight] Auto – The scale will turn on
when a button is pressed or the display weight
changes.
• Auto Off [A-Off] (on, off) – When Auto Off is set to
"ON," the scale will turn off automatically after five
minutes of inactivity. Auto Off is used to save battery
power.
• Filter [FILTER] (L1, L2, L3, L4) – Set the amount of signal
filtering.
Stability
Stabilization Time
• Auto Zero Tracking [AZT] (OFF, 0.5d, 1d, 3d, 5d, 8d,
10d) – Set the automatic zero tracking functionality.
The display will maintain zero until a change of
0.5d, 1d, 3d, 5d, 8d, or 10d divisions per second has
been exceeded.
• Stable [STAB] (0.5d, 1d, 2d, 5d) – Set the amount the
reading can vary while the stability symbol remains
on.
• Stable Compensation [STAB.C] (on, off) – Set the
automatic stable tracking functionality. Set it to "OFF"
for dosing or filing application.
L1
Less -------------------------------- Greater
Faster-------------------------------- Slower
L4
0422 IH-9884

Publicidad

loading

Este manual también es adecuado para:

H-9885H-9886