Descargar Imprimir esta página

Dusthunter T, Dusthunter C200, Fwe200Dh; Rated Data For Ambient Conditions; Rated Data For Devices With Ingress Protection; Supplements To Technical Data Section - SICK DUSTHUNTER C200 Instrucciones De Operación

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

10

DUSTHUNTER T, DUSTHUNTER C200, FWE200DH

10.1

Rated data for ambient conditions

Use inside or outside buildings
Altitude
Ambient temperature
Max. relative humidity
Line voltage fluctuations
Overvoltage category
Wet environment
Degree of contamination
[1] Recommendation: Use optional weather protection hoods for device components.
10.2

Rated data for devices with ingress protection

Enclosure rating
[1]
[1] Fitting position of MCU control unit Cables must be routed at the bottom to ensure IP protection
10.3

Supplements to Technical Data Section

10.3.1
Electrical data
8017305/AE00/V1-0/2022-09| SICK
Subject to change without notice
DUSTHUNTER T, DUSTHUNTER C200, FWE200DH
T50/100/200
[1]
Indoor and outdoor
Max. 2000 m (above sea level)
‒20 ... +60 °C
Option MCU-P: ‒40 ... +45 °C for purge
air intake temperature
<95% non-condensing
90 ... 250 V AC, 47 ... 63 Hz;
opt. 24 V DC, ±2 V
Option purge air unit SLV4 (2BH13):
3-phase: 200...240 V/345...415 V, 50 Hz
3-phase: 220...275 V/380...480 V, 60 Hz
Overvoltage category II
Not relevant
Degree of contamination 2
T50/100/200
IP66
Option purge air unit: IP54
Power connection of MCU control unit
Conductor
Power cable
Power connection of FWE200DH measuring and control unit
Conductor
Power cable
Battery
Battery (memory buffer)
Battery type:
Chemical system:
Device component:
A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | Dust measuring devices and Combined probe
C200
[1]
Indoor and outdoor
Max. 2000 m (above sea level)
‒40 ... +60 °C
Option MCU-P: ‒40 ... +45 °C for purge
air intake temperature
<95% non-condensing
90 ... 250 V AC, 47 ... 63 Hz;
opt. 24 V DC, ±2 V
Option purge air unit SLV4 (2BH13):
200...240 V/345...415 V, 50 Hz
220...275 V/380...480 V, 60 Hz
Overvoltage category II
Not relevant
Degree of contamination 2
C200
IP66
Option purge air unit: IP54
Cross-section
1.5 mm
2
... 4 mm
2
Cross-section
2.5 mm
2
... 6 mm
2
Button cell BR2032
Lithium poly-carbon monofluoride
MCU control unit
10
FWE200DH
[1]
Indoor and outdoor
Max. 2000 m (above sea level)
‒20 ... +50 °C
‒40 ... +45 °C for purge air intake
temperature
<95% non-condensing
90 ... 250 V AC, 47 ... 63 Hz
Overvoltage category II
Not relevant
Degree of contamination 2
FWE200DH
IP54
EN
25

Publicidad

loading