Descargar Imprimir esta página

Condenser Centering; Aperture Diaphragm - Optika Italy B-290 Serie Manual De Instrucciones

Ocultar thumbs Ver también para B-290 Serie:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 57
8.8

Condenser centering

The condenser is installed and pre-centered in the factory.
To remove the condenser use an Allen wrench 1.5 mm and
operate on the fixing knob placed on the right side of the
condenser holder.
Should a new centering is needed, operate in this way:
1. Insert 4x objective in the light path (in case 4x is not available
use the lower magnification available).
2. Focus the specimen.
3. Close the aperture diaphragm using the ring ①, moving the
ring to the value "4" related to the 4x objective. (Fig. 29)
4. Raise the condenser to the upper limit using the height ad-
justment knob ② placed on the left side of the condenser
holder.
5. Center the condenser using the centering screws ③ until the
field of view is evenly illuminated (in the field of view no dark
and bright areas must be noticed).
6. Fully open the diaphragm.
8.9

Aperture diaphragm

The Numerical Aperture (N.A.) value of the aperture dia-
phragm affects the image contrast. Increasing or reducing
this value one can vary resolution, contrast and depth of fo-
cus of the image. Move the diaphragm ring
the value corresponding to the objective in use. In this case
the optimal setting of the condenser is achieved.
With low contrast specimens set the numerical aperture to
about 70%-80% of the objective's N.A. If necessary, remove
on eyepiece and, looking into empty sleeve, adjust the con-
denser's diaphragm in order to obtain an image like the one
in Fig. 30.
APERTURE
DIAPHRAGM
(Fig. 29) on
Page 18
70-80%
FIELD OF VIEW
F ig. 29
F
ig. 29
30-20%
F ig. 30
F
ig. 30

Publicidad

loading