Please read carefully the
following instructions
and specifications be-
fore using the calibrator
and its accessories in
hazardous areas. Also
read the warnings avail-
able in Part A of MC6-Ex
User Manual.
Input Parameters, Simple Connections
Output Parameters, Simple Connections
Combined Connections
Connections are handled in groups, which partially share their protective circuitry. If using multiple meas-
urements/generations within one group simultaneously, the total power, capacitance and inductance shall
be within given limits for the whole group. For example, connections TC1 and TC2 belong to same group
with a maximum total power limit of 0.5 W. If the input power for connection TC1 is 0.4 W, the maximum
allowable input power for connection TC2 is 0.1 W when TC1 and TC2 are used simultaneously. Voltage
and current limit values apply to each connection, i.e. the voltage in both TC1 and TC2 may be 30 V, if
the total power does not exceed 0.5 W.
For other combined connections, all limit values apply for each connection. For example, 0.5 W can be
applied to TC1 at the same time as 0.5 W is applied to connection R1.

MC6-Ex Safety Information

The tables below refer to connections labelled in the picture below.
Connection
TC1, TC2
R1
R2
OUT
IN: V/Hz, Com
IN: mA, Com
HART/FF/PA, Com
SUPPLY, mA
Connection
TC1, TC2
R1
R2
OUT
IN: V/Hz, Com
IN: mA, Com
HART/FF/PA, Com
SUPPLY, mA
U
I
/V
/mA
i
i
30
215
30
215
30
215
30
215
30
215
24
215
Ui/V
Ii/mA
2.58
47.9
5.36
82.2
2.58
17.1
18.3
338.3
5.36
348.2
5.36
13.8
23.1
179.5
4.72
192.2
2
P
C
L
/mW
/nF
/mH
i
i
i
500
26.2
500
26.2
500
5.2
1000
31.3
1000
3.7
1000
7.3
0.01
Pi/mW
Ci/nF
L
/mH
i
30,8
39.8
15
102
39.8
11
60.8
100
482
34.7
0.1
18,1
62.3
100
1025
53.7
0.7
0
0
0
0
0
5