Surge immunity
Test
standard
EN 61000-4-5
Immunity to conducted disturbances induced by RF fields
Test
standard
EN 61000-4-6
en
Power frequency magnetic field immunity
Test
standard
EN 61000-4-8
34
The product must not be affected to initiate actions or to
render components defective as a result of lightning.
Immunity test level
Line-to-line: +/- 0.5 kV, +/-1 kV
Line-to-ground: +/- 0.5 kV, +/-1 kV, +/- 2 kV (N/A)
The wires must not be affected by other RF fields so as to
initiate actions or render components defective.
Immunity test level
3 V (6 V in ISM bands and amateur radio bands)
0,15 to 80 MHz, 80 % AM at 1 kHz
The product must not be affected to initiate actions or
to render components defective as a result of magnetic
fields.
30 A/m, 60 Hz