Voltage dips
The product must not initiate or limit actions as a result of
a dip in the power supply.
Test
Immunity test level
standard
EN 61000-4-11 0 % U
0 % U
70 % U
Single phase at 0°
Voltage interruptions
The product must not initiate or limit actions as a result of
a short-term power supply stop.
Test
Immunity test level
standard
EN 61000-4-11 0 % U
; 0.5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
T
; 1 cycle at 0°
T
; 25 cycles at 0°
T
; 300 cycle
T
en
35