S
ETTINGS
S
S
TANDARD
ETTINGS
HORIZON 6FA
RegenLab
R
C
T
EVIEW
YCLE
IME AND
Factory programmed cycles are shown on the rear of the centrifuge, on the Factory Set Cycles label. To review
current settings, follow this procedure:
o The lid must be open to review the selected cycle time and speed.
o Press and hold the START button until you hear a beep.
o Release the START button. The centrifuge will beep and the LED light will flash once for each minute of run time
in the current cycle. 10 beeps / flashes equal 10 minutes of run time. Run time starts when the rotor reaches 90%
of desired speed and stops when the rotor starts decelerating.
o Pressing the START button again will cause the unit to beep and the LED light to flash once for each 100 rpm in
the current cycle. 38 beeps / flashes equal 38 x 100 or 3,800 Revolutions per Minutes (RPM)
o The centrifuge will automatically revert to normal mode at the end.
C
C
T
HANGING
YCLE
IME AND
o The lid must be open to change the selected cycle time and speed.
o Select the cycle you wish to change.
o Press and hold the START and UNLOCK buttons together until the LED light flashes.
o Press the START button for each minute of run time.
o Move to speed setting mode by pressing the UNLOCK button.
o Press the START button once for each 100 rpm.
o Press the UNLOCK button to exit setting mode.
(1) 5 min
3500 RPM
RPM
5 minutes
Time
1500xg
G-Force
S
S
PEED
ETTINGS
S
S
PEED
ETTINGS
RegenLab SA – Customer Service: +41 (0)21 864 0111 | Page 5
(2) 9 min
(3) 30 min
3500 RPM
3500 RPM
9 minutes
30 minutes
1500xg
1500xg