LED SURGICAL HANDPIECE WITH EXTERNAL IRRIGATION
CAUTIONS
Please read this instruction before use and file for future reference.
Prior to each use, all instruments and accessories must be inspected for proper operation.
When using this product, pay attention to protect the patient at all time.
This product should not be rebuilt. Replacing bearings only could result in unexpected failure and patient injury due to di erent factors of new vs. old components.
End of product life, the metal can be handled by separation and recycling disposal.
P roducts storage and transportation should be carried out in environment with the following features: dry, stable (no collision), clean, normal temperature, normal
pressure, away from acid gases and other harmful chemical substances, and humidity should be less than 90%.
Maintenance should be made by professionals.
This product should be used with caution for the patients with bleeder disease, pacemaker, and pregnant woman or children.
12
1. Product description
en
This medical device is:
– intended for dental treatment only. All other types of use or alterations to the product are not permitted and can be hazardous. The medical product is
– designed to be used with the corresponding heads in the following applications: surgery e.g. bone augmentation, sinus lift, dental extraction, implantology and
maxillo-facial surgery.
– A medical device according to relevant national statutory regulations. This product can be connected to motors with a connection in accordance with EN ISO 3964.
During the operation, Sterile physiological coolant is required, and water spray on the unit must be turned o .
E type lubrication tip
Fig. 1