• Each legrest separately on the spreader bar
L
CAUTION: Risk of falling- This method increases the chance of sliding out the
sling.
Attaching each legrest separately on the spreader bar can be used for patient with sensitive
genitals. The attachment of the sling with this method shall be according instructions below:
1. Put the legrest around each leg separately, but not crossed.
2. Attach the mounting straps of the legrests on the spreader bar of the patient hoist.
3.2.2 Position of spreader bar
The spreader bar must always be perpendicular to the body of the patient. The left figure is the
good method. The right figure is the wrong way.
3.3
Battery and charging
3.3.1 Battery charger
Primary voltage
Safeguards
Ambient temperature
Length of mains lead
Ambient storage temperature
Relative storage humidity
Conformity
3.3.2 Charging the batteries
Use only the control box with included battery charger.
We recommend that you should regularly recharge the batteries in order to ensure that the hoist
remains usable and to extend the lifespan of the batteries. The control box will give a warning
sound to tell you when the battery charge is low.
First use
First, insert the supply cable with the connector side into the appropriate connector on the control
box. The plug side of the supply cable must be insert in the socket outlet. The loading time is
about 24 hours.
Recharging
WARNING: Risk of injury – Do not use the lift when the supply cable is connected in
L
the socket outlet.
When the battery is running low, only the left LED will light green. Once the battery level gets
very low, the LED will light red and a beeping sound can be heard every 2 seconds.
100 - 240VAC / 50/60 Hz
Protected against reverse polarisation, electrical surges
and extreme temperature
0 to +40°C
2.0 m
-15° to +50°C
max. 95% (non-condensing)
tested in accordance with IEC 60601-1
9
EN
Falcon
2022-11