Recharging the Battery (For FORA D40g Only)
The power adapter can be used as the battery charger when you need to recharge Li-ion
battery. To recharge the 3.7V Li-ion battery, connect AC adapter to the monitor and a power
source.
The power indicator will light up in green, which indicates the battery is undergoing a
recharge. It should take approximately 5 hours to be fully charged.
If the power indicator starts blinking in green, this indicates the recharge is nearly
finished. The power indicator will diminish when the recharge completes.
Using an AC Adapter
As an alternative, you can use an AC adapter to provide the power supply. However, when
alkaline batteries are used for FORA D40a and FORA D40b, these batteries cannot be
recharged with the device by connecting AC adapter.
To use an AC adapter, do the following:
1. Connect AC adapter plug to AC adapter jack of the device.
2. Plug AC adapter power plug into an electrical outlet.
CAUTION!
• For D40g, the Li-ion battery must be recharged with the FORA D40g monitor by connecting
AC adapter to an electrical outlet. Only this 3.7V rechargeable Li-ion battery (battery pack no.
S14500 1S4P) can be recharged with FORA D40 by the connection of AC adapter. Any other kind
of rechargeable battery is not allowed or it may damage the monitor.
Caring for Your Device
To clean the device exterior, wipe it with a cloth moistened with tap water or a mild cleaning
agent, then dry the device with a soft dry cloth. Do NOT flush with water.
• Do NOT use organic solvents to clean the device.
• Do NOT wash or iron the pressure cuff.
• Always store or transport the device in its original storage case.
• Avoid dropping and heavy impact.
• Avoid direct sunlight and high humidity.
Device Storage
• Storage condition: -20°C to 60°C (-4°F to 140°F), below 95% relative humidity.
• Always store or transport the monitor in its original storage case.
• Avoid dropping and heavy impact.
• Avoid direct sunlight and high humidity.
Caring for Your Test Strips
• Storage condition: 4°C to 40°C (39.2°F to 104°F), below 85% relative humidity. Do not freeze.
• Store your test strips in their original vial only. Do not transfer to other container.
• Store test strip packages in a cool and dry place. Keep away from direct sunlight and heat.
• After removing a test strip from the vial, immediately close the vial cap tightly.
• Touch the test strip with clean and dry hands.
• Use each test strip immediately after removing it from the vial.
EN-20
D40 Owner's Manual