Emissivity Values
Substance
Asphalt
Concrete
Cement
Sand
Earth
Water
Ice
Snow
Glass
Ceramic
Marble
Plaster
Mortar
Brick
6

Measurement operation

Hold the meter by its Handle Grip and point it toward the sur-
face to be measured.
Pull and hold the Trigger to turn the meter on and begin test-
ing. The display will light if the battery is good. Replace the
battery if the display does not light.
While measuring, the SCAN display icon will appear in the up-
per left corner of the LCD.
Release the Trigger and the HOLD display icon will appear on
the LCD indicating that the reading is being held. In HOLD sta-
tus, press the Laser/Backlight button to turn on or off the laser
and the backlight.
The meter will automatically power down after approximately 7
seconds after the Trigger is released. (Unless the unit is
locked on.)
Thermal
emissivity
0,90 to 0,98
0,94
0,96
0,90
0,92 to 0,96
0,92 to 0,96
0,96 to 0,98
0,83
0,90 to 0,95
0,90 to 0,94
0,94
0,80 to 0,90
0,89 to 0,91
0,93 to 0,96
DIT-130 – USER MANUAL
Substance
Cloth (black)
Human skin
Leather
Charcoal (powder)
Lacquer
Lacquer (matt)
Rubber (black)
Plastic
Timber
Paper
Chromium oxides
Copper oxides
Iron oxides
Textiles
Thermal
emissivity
0,98
0,98
0,75 to 0,80
0,96
0,80 to 0,95
0,97
0,94
0,85 to 0,95
0,90
0,70 to 0,94
0,81
0,78
0,78 to 0,82
0,90
35