3. Instructions for use
The device is designed to measure the temperature and pH / ORP / EC of test
solutions.
The user is responsible for any damage resulting from misuse.

3.1. Product overview

1 - LCD display
2 - Power switch
3 - Probe cover
4 - Battery compartment cover
5 - HOLD button
6 - Unit change button (°C/F)

3.2. Working with the unit

When the probe tip is immersed in the test solution, air bubbles can settle on its
surface, leading to errors in readings. Therefore, when measuring and calibrating
the device, it is recommended to slowly move the probe in a circular motion to
minimize the occurrence of this issue.
If too large measurement deviations occur with a properly performed calibration,
the measurement results do not reset, etc., clean the electrode with ethyl alcohol
and check the battery charge level.
PERFORMING MEASUREMENTS
7 - Unit change / calibration button 8 -
Calibration trimmer
9 - pH 7 calibration trimmer
10 - pH 4 calibration trimmer
11 - Calibration trimmer
12 - Maximum immersion level