E E x x p p l l a a n n a a t t i i o o n n O O f f S S y y m m b b o o l l s s : :
OFF/ON:__________________________________________________________
Class II Equipment:_________________________________________________
Type B Applied Part:__________________________________________________
Ordinary Equipment (IPX0) According to the Ingress of Water.
Mode of Operation: _____________________________________________continuous
T T h h e e F F o o l l l l o o w w i i n n g g T T a a b b l l e e s s A A r r e e P P r r o o v v i i d d e e d d I I n n A A c c c c o o r r d d a a n n c c e e W W i i t t h h I I E E C C 6 6 0 0 6 6 0 0 1 1 - - 1 1 - - 2 2 : :
Guidance and manufacturer's declaration –electromagnetic emissions
®
The PRONEB
ULTRA II is intended for use in the electromagnetic environment specified below. The customer or the
®
user of the PRONEB
ULTRA II should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
English
7
Compliance
Electromagnetic environment – guidance
®
The PRONEB
ULTRA II uses RF energy only for its
internal function. Therefore, its RF emissions are very low
Group 1
and are not likely to cause any interference in nearby
electronic equipment.
Class B
®
The PRONEB
ULTRA II is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
Not applicable
supply network that supplies buildings used for domestic
purposes.
Not applicable