6
Performance Tests and Calibration
For serial numbers below MY51530001, the 10 kHz input frequency is applied to
NOTE
those marked with an asterix (*)
Table 6-4
Adjustment table
Step
Test Function
1
Turn the rotary switch to the
position
2
Turn the rotary switch to
position (for model
U1252B), to
position (for model U1251B)
170
Cal Range
Input
5 V
0.3 V,1 kHz
3 V, 1 kHz
3 V, 20 kHz *
50 V
3 V, 1 kHz
30 V, 1 kHz
30 V, 20 kHz *
500 V
30 V,1 kHz
300 V,1 kHz
300 V, 20 kHz *
1000 V
30 V, 1 kHz
300 V, 1 kHz
300 V, 20 kHz *
Short
Dual Banana Plug
with copper wire
short between 2
terminals
5 V
3 V
50 V
30 V
500 V
300 V
1000 V
1000 V
Keysight U1251B and U1252B User's and Service Guide
Cal Item
U1251B
U1252B
0.3000 V
0.3000 V
3.0000 V
3.0000 V
3.0000 V
3.0000 V
03.000 V
03.000 V
30.000 V
30.000 V
3.0000 V
30.000 V
030.00 V
030.00 V
300.00 V
300.00 V
3.0000 V
300.00 V
0030.0 V
0030.0 V
0300.0 V
0300.0 V
3.0000 V
0300.0 V
SHort
SHort
3.0000 V
3.0000 V
30.000 V
30.000 V
300.00 V
300.00 V
1000.0 V
1000.0 V