Display Disappear
• Check whether the low battery icon is appeared
before the display is off. If yes, replace with new
batteries
Air Bubbles Adhere On Electrode
• Stir the electrode completely and better to dip
the electrode into solution at oblique angle. After
soaking the electrode for 15~30 minutes, inspect
the electrode carefully to make sure no bubbles
adhere.
• If air bubbles still exist, tap the bottom of the
container gently and stir the electrode to remove
the air bubbles. If above method are not working,
remove the electrode out of solution and blow at
the electrode to remove the air bubbles
Error Code
Parameter: Conductivity
"- - -"
Meter is in manual range 1 however the
measured value is higher than 1999uS.
• Press "p" more than 2 seconds to change
the range to range 2 or to auto ranging.
"E03"
Conductivity value is over the range limit
(19.99mS) or meter is damaged.
• Put the meter in standard solution. If E03
still appears, send back for repair.
"E04"
Caused by temp. reading error
• Refer to error code of temp. After solving
the error of temp, E04 will disappear.
13