The ASPEED 3.0 aspirator is a professional device suitable for home and outpatient use, specific for secretion-aspiration. It has a vacuum
regulator, a vacuum gauge, a 1000 ml sterilisable vessel, or (depending on the version) a 2000 ml sterilisable vessel; these vessels are equipped
with a protective device to prevent liquid from entering the suction pump and interrupting the aspiration flow. It is lifetime lubricated, easy to
handle, simple to use, reliable, strong and silent.
The ASPEED 3.0 aspirator, in all its versions, has the following accessories: a 1000 ml vessel or 2000 ml vessel (depending on the version)
with protection device, 34 cm long sterilisable silicone connection tube (only for 1000 ml vessel version), no. 2 20 cm long sterilisable silicone
connection tubes (only for 2000 ml vessel version), 130 cm sterilisable silicone connection tube, sterile disposable cannula, sterile dis-
posable manual flow regulator, disposable bag (only for 1000 ml vessel version) and disposable antibacterial filter.
N.B.: Only use genuine accessories supplied by 3A Health Care; the accessories supplied with the device by the
manufacturer have been tested and proven compliant with the latest current safety standards. WARNING! Correct
operation of the device is not guaranteed in the event of use of accessories other than those supplied.
Mains-operated medical /surgical aspirator intended for use in home care and/or outpatient settings.
Medical purposes: This product is intended for use for the aspiration of body fluids.
Intended users of the product:
• Legally certified medical personnel (doctors, nurses and therapists).
• For home treatment or home care under the guidance of medical personnel.
• Users must also be able to understand the operation of the medical device, and the contents of the instruction manual, in general terms.
Intended patients for the product: People who need to remove body fluids (saliva, blood, etc.).
Environment: This product is intended for use in home care and/or outpatient settings.
Expected duration: Duration may vary based on the operating environment. The lifetime of the device is 5 years and that of the
collection vessel and the silicone tubes 1 year or 30 sterilisation cycles. The cannula, manual flow regulator, antibacterial filter and bag
are disposable devices and must therefore be replaced after each application. Frequent use of the product may shorten the duration.
Precautions for use: The warnings and cautions described in the instruction manual must be observed.
This is a medical device and it must be used only upon medical prescription. It must be used as described in this instruc-
tion manual. The user must read and understand in full the information regarding the use and maintenance of the unit. For
any question, do not hesitate to contact your preferred retailer.
MICROBIAL CONTAMINATION: due to pathologies with risk of microbial infection and contamination a personal use of the
accessories is recommended (ask your physician).
The manufacturer does anything possible to provide a product equipped with the highest quality and safety features, although, being it an
electric device, some fundamental safety directions must be observed:
• Kids and non self-sufficient people must always use the device under strict supervision of a competent adult in his full-capacity and who has
read and understood this manual. Ensure that children do not use the device unsupervised some parts are small enough to be swallowed.
• The device must always be used by specifically trained staff who have read this manual.
• This device must only be used for its intended purpose: aspirator for outpatient and domestic use; any other use is considered improper and
dangerous and the manufacturer cannot be held responsible for any consequence of misuse.
• Maintenance and/or repair must be carried out only by the customer care service. Non-authorised repairs void the warranty.
• Never use adapters for voltages different from the voltage indicated on the data label on the bottom of the device.
• Keep the cable away from hot surfaces.
• This device is not suitable for use in presence of anaesthetic blend inflammable with air, oxygen, or nitrous oxide.
Never handle the power supply cable plug with wet hands, and never use the device in the bath or in the shower.
• Never leave the device near water, never immerse it in any liquid, never wet it, if it gets wet immediately unplug it from the power supply socket
before touching it. Do not use the device if the plug or the power supply cable are worn or wet; immediately send it to the customer care
service. In case of damage to the power supply cable, this should only be replaced by the customer care service.
EN
INTENDED USE
IMPORTANT WARNINGS