3.8 EMC Table
Radiated RF Emission levels
Phenomenon
Emissions
Radiated RF Emissions
CISPR 11
Radiated RF Emissions
CISPR 11
Immunity levels
Immunity test levels
Phenom-
Professional
enon
healthcare facility
environment
Immunity
Test
Electrostatic
Discharge
IEC
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
61000-4-2
Radiated
3 V/m
Immunity
80 MHz-2,7 GHz
IEC 61000-
80% AM at 1 kHz
4-3
Power
Frequency
Magnetic
Field
IEC 61000-
4-8
1
Complies without any deviations from IEC 60601-1-2
and normative references.
Professional healthcare facility
environment
Home
healthcare
environment
1
Compliance
± 8 kV contact
10 V/m
80 MHz-2,7 GHz
80% AM at 1 kHz
30 A/m
50 Hz or 60 Hz
Immunity test levels
Home healthcare environment
Compliance
Group 1
Class B
Proximity fields from RF wireless communication
equipment, Immunity test levels:
Frequency Range and Level: RF wireless
communication equipment
Frequency
(MHz)
Test
Pulse Modulation: 18Hz
385
450
FM + 5Hz deviation: 1kHz sine
710
745
Pulse Modulation: 217Hz
780
810
870
Pulse Modulation: 18Hz
930
1720
1845
Pulse Modulation: 217Hz
1970
2450
Pulse Modulation: 217Hz
5240
5500
Pulse Modulation: 217Hz
5785
2 Compliance with immunity test levels specified in IEC
60601-1-2, without deviations.
11
Modulation
2
Compliance
Immunity
Level
(V/m)
27
28
9
28
28
28
9

Publicidad