2.2 Home Cleaning & Disinfection
Home Cleaning
Clean Sunset Heated CPAP Tube before first use and daily during use.
1. Disconnect Heated CPAP Tube from CPAP machine and mask.
2. Thoroughly rinse inside and outside of Heated CPAP Tube with cold
tap water to remove excess soil.
3. Gently wash Heated CPAP Tube and connectors by immersing in
a solution of warm water and mild unscented liquid dish soap,
gently agitating by hand and using a soft bristle brush to remove
contaminants.
4. Thoroughly rinse Heated CPAP Tube with clean water to remove all
soap residue.
5. With connector cuffs directed down, toward the floor, gently tap
connector cuffs to remove excess water.
6. Hang Heated CPAP Tube with connector cuffs directed downward,
towards the floor, for drying.
7. Allow Heated CPAP Tube to air dry thoroughly prior to next use.
8. Visually inspect Heated CPAP Tube for cleanliness. Repeat cleaning
process if not visually clean.
9. Inspect Heated CPAP Tube for damage or wear. Discard and replace if
necessary.
Caution: Do not use harsh soaps, detergents, solvents, or alcohols to
clean the Sunset Heated CPAP Tube. Do not attempt to clean Heated CPAP
Tube in a dishwasher, ultrasonic cleaner or any other cleaning device.
Only reuse the Heated CPAP Tube when fully dry.
Home Disinfection
The disinfection procedure described is recommended for a maximum of
60 cycles.
Thermal Disinfection (Pasteurization)
Immerse Heated CPAP Tube in 75°C ± 2°C (167°F) water bath for 30
minutes.
Post Disinfection
Repeat steps 5-9 of the Home Cleaning procedure prior to use.